Stability / Humidity Chamber

CONSTRUCTION

  • Double wall, outer Chamber made by cabinet mild steel powder coated or stainless steel and Inner Chamber made by Stainless Steel.
  • Full view glass door a multi pane viewing door prevents condensation & allows product view. Inner & outer doors are sealed around the entire perimeter by silicone sponge rubber gasket.
  • The chamber is insulated by filling PUF in the space between outer body and chamber to prevent temperature loss.(Insulation-75mm)
  • Chamber illumination: LED light with door switch.
  • Castor wheels for easy move ability
  • Trays: Stainless steel.
 

TECHNICAL DATA

    • For conditions as per ICH guidelines ( International Conference of Harmonization) 25°C – 60% RH/40°C – 75% RH/ 30°C – 65% RH/25°C – 40% RH/ 40°C – 25% RH/ 30°C – 35% RH.
    • Temperature Range : 10.0°C to 60.0°C
    • Accuracy/ Uniformity : ± 1°C / ± 2°C
    • Humidity Range : 25% RH to 95% RH
    • Accuracy/ Uniformity : ± 3 % RH / ± 5 % RH or better
    • Temperature / Humidity display : Digital LED.
    • Digital Temp. Controller : Microproceser based auto Tune  P.I.D Digital Temperature. Controller  with PT 100 Sensor. (Make: i-therm Or equivalent)
    • Resolution : 0.1°C / 0.1% RH
    • Temperature sensor : PT – 100
    • RH sensor : 0 to 3.3 V DC
    • Heating : ‘U’ shaped tubular heater in stainless steel.
    • Cooling : CFC free compressor utilizing R 134A eco friendly refrigerant, with condenser, motor and relay or as per the requirement of customer for higher accuracy.(Compressor make: Emerson or equivalent)
    • Steam injection : Boiler with reservoir made of thick stainless steel, with heater, water inlet control, low water level safeguard
    • Air circulation: cooling type fan with impeller / blower and flange type motor.
    • Electrical: 230V / 15A / 50 Hz

DOCUMENTATION

  • We provide Manual &DQ, IQ, OQ, PQ documents compliant to competitive regulatory guidelines.
  • We provide calibration certificates & Temperature and R.H Mapping Report for Stability Chamber, which
  • are traceable to national standards.
  • We also provide S.O.P for Stability Chamber.

SAFETY FEATURES

  • High Temperature safety cut off, Low water level boiler cut off
  • Electrical overload cutoff relay for compressor, Time delay for compressor switch ON

PRE INSTALLATION MANDATORY REQUIREMENT

  • Distilled water or DM water continuous supply at 3 feet height.
  • Room temperature around machine preferably at 25°C with air conditioning or a well-ventilated room With exhaust fan. However, surrounding temperature should not exceed 30°C.
  • Stabilized input voltage of 230V A.C Use of servo controlled stabilizer is recommended.
  • Water Drain line with 1/2″ nozzle to be provided at floor level.

OPTIONAL ACCESSORIES

  • STAND BY REFRIGERATION SYSTEM
  • Stand by compressor manual switchover facility with compressor time delay
  • PC/ PRINTER INTERFACE UNIT
  • Facility to print records on printer with set values, actual values, date and time with variable print time
  • SOFTWARE
  • 21-CFR PART 11 COMPLIANCE SOFTWARE
Model No. Size (WxDxH) Liters No.of Shelves K.W
JDJ/SC-01 18″ X 16″ X 20″ 90 2 2.5
JDJ/SC-02 20’’ x 20’’ x 28’’ 175 2 2.5
JDJ/SC-03 22″ x 22″ x 36″ 270 3 3.0
JDJ/SC-04 24″ x 24″ x 36″ 320 3 3.0
JDJ/SC-05 24″ x 24″ x 48″ 440 43.5

* Customize size provide on user Request.